2 Fluke Corporation 719Pro Electric Pressure Calibrator
Pressure measurement Range Resolution Accuracy Comment
Fluke-719Pro-30G -12 to 30 PSI
/-0.8 to 2 bar
0.001 PSI,
0.0001 bar
± 0.025 % of full scale, 6 months
± 0.035 % of full scale, 1 year Dry Air and non-corrosive gas onlyFluke-719Pro-150G -12 to 150 PSI
/-0.8 to 10 bar
0.01 PSI,
0.001 bar
Fluke-719Pro-300G -12 to 300 PSI /
-0.8 to 20 bar
0.01 PSI,
0.001 bar
Temperature effect
(all ranges)
Add ± 0.002 % F.S./°C for temps
outside of 15 °C to 35 °C
No effect on accuracy on all functions
from 15 °C to 35 °C
Function Range Resolution Accuracy (1 year) Comment
mA dc (simulation and
measurement)
0-24 mA dc 0.001 mA ± 0.015 % of reading ± 2 counts 1000 ohm maximum load in mA source,
26 V dc maximum voltage in mA simulate
Volts dc (measurement only) 30 V dc 0.001 V ± 0.015 % of reading ± 2 counts No AC voltage measurement, do not
exceed 30 V dc
Temperature (measurement
only, Pt100 RTD)
-50 to 150 °C
(-50 °F to 302 °F)
0.01 °C
0.01 °F
Temperature ± 0.1 °C (0.2 °F) ± 0.25 °C (± 0.45 °F) combined
uncertainty when using 720 RTD probe
(optional accessory)
Loop power supply 24 V dc N/A 24 mA at 24 V
Temperature effect
(all functions)
Add ± 0.002 % F.S./°C for temps
outside of 15 °C to 35 °C
No effect on accuracy on all functions
from 15 °C to 35 °C
Environmental
Operating temperature -10 °C to +50 °C
Storage with batteries per battery manufacturer’s specification within storage specification below
Storage temperature (without batteries) -20 °C to +60 °C
Power requirements
Battery Eight (8) standard Alkaline AA batteries
Battery life 300 pump cycles to 150 PSI minimum; 1000 pump cycles to 30 PSI;
100 pump cycles to 300 PSI minimum
Physical
Dimensions (HxWxD) 24.13 cm x 11.18 cm x 7 cm (9.5 in x 4.4 in x 2.8 in)
Weight With holster 1.2 kg (2.5 lb)
Connections
Pressure 1/8 inch female NPT
Electrical Standard banana jacks
RTD 4 pin miniature circular
External module 6 pin miniature circular
Engineering units (configurable)
psi, bar, mbar, kPa, kgcm2, cmH2O@4 °C, cmH2O@20 °C, mH2O@4 °C, mH2O@20 °C, inH2O@4 °C, inH2O@20 °C, inH2O@60 °F, mmHg@0 °C,
inHg@0 °C, ftH2O@4 °C, ftH2O@20 °C, ftH2O@60 °F
Specifications
Fluke Corporation
PO Box 9090, Everett, WA 98206 U.S.A.
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or
Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2013 Fluke Corporation.
Specifications subject to change without notice.
Printed in U.S.A. 7/2013 4347338A_EN
Modification of this document is not permitted
without written permission from Fluke Corporation.
Fluke. The Most Trusted Tools
in the World.
Ordering information
Fluke-719Pro-30G Electric Pressure Calibrator, 30 PSIG
Flu ke -719Pro-150G Electric Pressure Calibrator, 150 PSIG
Fluke-719Pro-300G Electric Pressure Calibrator, 300 PSIG
Fluke-720RTD RTD Probe for 721 and 719Pro
Fluke-720URTDA Universal RTD adapter for Fluke-721 and 719Pro
Fluke-71XTrap Liquid and dirt trap for 718 and 719Pro
Included with all models
Test leads with alligator clips, printed users manual and users manual
on CD-ROM in 14 languages, fluid trap, carrying case, 1/4 inch NPT
to ¼ inch ISO adapter, accessory hose kit with (2) 1/8 inch quick fit
adapters and 1 meter of hose, Traceable Certificate of Calibration.