Agilent U3606A Multimeter|DC Power Supply Data Sheet The real two instruments in one box Get to know the U3606A's replacement - the U3606B www.agilent.com/find/dmm-power Features * Two independent instruments in one box * Nine basic measurements as well as 4-wire milliohm measurement and eight math functions * Measurement speed: up to 37 readings/s * Low error rate: up to 0.025% DCV accuracy * 30-W dual-range power supply with remote sensing * Excellent load regulation: up to 0.01%+3 mV * OVP and OCP load protection * Auto scan and ramp, and built-in 4.8 kHz square-wave generator * USB-TMC488.2 and GPIB connectivity * Kensington lock slot security Pick both Digital multimeter (DMM) and power supply? Simultaneous and independent? Small and cost-effective? Get all this and more in the Agilent U3606A multimeter|DC power supply. This convenient new hybrid combines a 51/2-digit DMM and 30-W dual-range supply in a single unit. Operating simultaneously and independently, the instruments provide efficient, affordable testing while saving space on the bench or in a rack. The 5.5-digit DMM The 5.5-digit DMM includes nine essential multimeter capabilities as well as 4-wire milliohm measurement and eight built-in math functions. The DMM also provides a fast measurement speed of up to 37 readings/s, and a low error rate of up to 0.025% DCV accuracy. Physical security and seamless system integration Instruments may be at risk of theft or misplacement when left unattended on the bench. With the hybrid multimeter's rear Kensington lock slot, you can secure your instrument and be assured that it is where you expect it to be for your continued testing the next day. The rackmountable U3606A also enables seamless integration into your system via popular GPIB and USB-TMC488.2 interfaces, programmable with standard SCPI commands. The 30-W DC power supply The 30-W DC power supply provides a dual-range output of 30 V/1 A and 8 V/3 A, with an excellent load regulation of up to 0.01%+3 mV. The power supply adds overvoltage and overcurrent load protection (OVP and OCP), a built-in square-wave generator, and auto scan and ramp for multilevel DC bias testing. Remote sensing capability further ensures accurate supply of power at load end. Square-wave generator Sweep functions Square-wave output is a unique function for many applications, such as pulse-width modulation (PWM) output, adjustable voltage control, and synchronous clock (baud rate generator). You can also use this function to check and calibrate flowmeter displays, counters, tachometers, oscilloscopes, frequency converters, frequency transmitters, and other frequency input devices. The U3606A's square-wave output provides selectable frequencies up to 4.8 kHz with variable duty cycles and amplitudes. Sweep functions in the U3606A are auto ramp and scan outputs for lowspeed multilevel DC bias tests, such as margin tests, power cycling tests and relay control. Both functions are conveniently configurable from the front panel to sweep up to 100 steps for scan and 10,000 steps for ramp, programmable to 105% of full scale. Auto scan output Amplitude (V/A) Amplitude end position Number of steps Time (t) Scan dwelling time Total dwelling time Auto ramp output Amplitude (V/A) Amplitude end position Number of steps Time (t) Total dwelling time 2 Take a closer look VFD display Source terminals Multimeter terminals Local/Shift key Power On/Off button Autorange and manual range keys Multimeter operation keys Source operation keys Figure 1. Front panel of the U3606A. Ventilation fan Kensington lock slot GPIB (IEEE-488) connector AC line fuse AC power connector USB 2.0 Type B connector Chassis ground lug Rear output terminals for remote sense Figure 2. Rear panel of the U3606A. 3 Digital multimeter specifications Specification assumptions: * * * * * * Specifications stated are after 60-minutes of warm-up and for 51/2-digit resolution One-year calibration cycle, with calibration temperature of 18 C to 20 C Operating temperature: 18 C to 28 C (64.4 F to 82.4 F) Accuracy is expressed as (% of reading + % of range) Temperature coefficient: Add [0.1 x (the specified accuracy) / C] for 0 C to 18 C and 28 C to 55 C Relative humidity (RH) up to 80% at 30 C, proportional to 50% for 30 C to 55 C DC specifications Temperature Accuracy (% of reading + % of range) coefficient 90 days 1 year 24 hours2 0 C to 18 C 23 C 1 C 23 C 5 C 23 C 5 C 28 C to 55 C Function Range1 Test current or burden voltage DC voltage 100.000 mV 1.00000 V 10.0000 V 100.000 V 1000.00 V - - - - - 0.012 + 0.008 0.012 + 0.005 0.012 + 0.005 0.012 + 0.005 0.012 + 0.005 0.015 + 0.008 0.015 + 0.005 0.015 + 0.005 0.015 + 0.005 0.015 + 0.005 0.025 + 0.008 0.025 + 0.005 0.025 + 0.005 0.025 + 0.005 0.025 + 0.005 0.0015 + 0.0005 0.0010 + 0.0005 0.0020 + 0.0005 0.0015 + 0.0005 0.0015 + 0.0005 DC current3 10.0000 mA 100.000 mA 1.00000 A 3.0000 A < 0.2 V < 0.2 V < 0.3 V < 0.7 V 0.05 + 0.015 0.05 + 0.005 0.05 + 0.007 0.05 + 0.007 0.05 + 0.015 0.05 + 0.005 0.05 + 0.007 0.05 + 0.007 0.05 + 0.015 0.05 + 0.005 0.15 + 0.007 0.15 + 0.007 0.0060 + 0.0005 0.0060 + 0.0005 0.0100 + 0.0005 0.0150 + 0.0010 Resistance4 100.000 1000.00 10.0000 k 100.000 k 1.00000 M 10.0000 M 100.000 M 0.83 mA 0.83 mA 100 A 10 A 900 nA 205 nA 205 nA || 10 M 0.04 + 0.008 0.04 + 0.005 0.04 + 0.005 0.04 + 0.005 0.05 + 0.005 0.20 + 0.005 1.60 + 0.005 0.04 + 0.008 0.04 + 0.005 0.04 + 0.005 0.04 + 0.005 0.05 + 0.005 0.20 + 0.005 1.60 + 0.005 0.05 + 0.008 0.05 + 0.005 0.05 + 0.005 0.05 + 0.005 0.06 + 0.005 0.25 + 0.005 2.00 + 0.005 0.0050 + 0.0005 0.0050 + 0.0005 0.0050 + 0.0005 0.0050 + 0.0005 0.0050 + 0.0005 0.0150 + 0.0005 0.1500 + 0.0005 Continuity 1.0000 k 0.83 mA 0.04 + 0.005 0.04 + 0.005 0.05 + 0.005 0.0050 + 0.0005 Diode5 1.0000 V 0.83 mA 0.04 + 0.005 0.04 + 0.005 0.05 + 0.005 0.0050 + 0.0005 Capacitance6 1.000 nF 10.00 nF 100.00 nF 1.000 F 10.00 F 100.0 F 1000 F 10000 F 0.75 A 0.75 A 8.3 A 83 A 83 A 83 A 0.83 mA 0.83 mA - - - - - - - - - - - - - - - - 2.0 + 0.8 1.0 + 0.5 1.0 + 0.5 1.0 + 0.5 1.0 + 0.5 1.0 + 0.5 1.0 + 0.5 2.0 + 0.5 0.02 + 0.001 0.02 + 0.001 0.02 + 0.001 0.02 + 0.001 0.02 + 0.001 0.02 + 0.001 0.02 + 0.001 0.02 + 0.001 1. 20% over-range on all ranges, except for 1000 Vdc range. 2. Relative to calibration standards. 3. Any current measurement greater than 500 mA will have a temporary thermo-effect. If you wish to measure a lower current or offset current immediately after a high-current measurement, ensure that the U3606A has cooled down. 4. Specifications stated are for 2-wire resistance measurements using Null math operation. Without Null, add a 0.2 error. To eliminate noise interference which may be induced by the test leads, a shielded test cable is recommended for resistances above 100 k. 5. Specifications stated are for the voltage measured at the input terminals only. The test current (1 mA) is typical. Variation in the current source will create some variation in the voltage dropped across a diode junction. 6. Specifications stated are for open test lead measurements with film capacitor or better, using Null math operation. 4 Digital multimeter specifications (continued) Low-resistance specifications Range Test current Accuracy (% of reading + % of range)1 1 year (23 C 5 C) 100 m 1.0000 A 0.25 + 0.05 1000 m 0.1000 A 0.25 + 0.03 10 0.1000 A 0.25 + 0.03 1. Four-wire measurement method is used. Test current is sent from the FORCE terminals and resistance is measured at the SENSE terminals. AC specifications Frequency range Temperature Accuracy (% of reading + % of range) coefficient 90 days 1 year 24 hours2 0 C to 18 C 23 C 1 C 23 C 5 C 23 C 5 C 28 C to 55 C 100.000 mV to 750.00 V4 20 Hz to 45 Hz 45 Hz to 10 kHz 10 kHz to 30 kHz 30 kHz to 100 kHz7 0.60 + 0.1 0.16 + 0.1 0.80 + 0.1 3.00 + 0.2 0.60 + 0.1 0.16 + 0.1 0.80 + 0.1 3.00 + 0.2 1.00 + 0.15 0.20 + 0.1 1.00 + 0.16 3.00 + 0.28, 9 0.02 + 0.02 0.02 + 0.02 0.02 + 0.02 0.05 + 0.02 10.0000 mA to 3.0000 A10 20 Hz to 45 Hz 45 Hz to 1 kHz 1 kHz to 10 kHz11 0.80 + 0.1 0.40 + 0.1 2.00 + 0.2 0.80 + 0.1 0.40 + 0.1 2.00 + 0.2 1.50 + 0.1 0.50 + 0.1 2.00 + 0.2 0.02 + 0.02 0.02 + 0.02 0.02 + 0.02 Function Range1 True rms AC voltage3 True rms AC current 1. 20% over-range on all ranges, except for 750 Vac range. 2. Relative to calibration standards. 3. Specifications stated are for input signals greater than 5% of range. 4. Available ranges: 100.000 mV, 1.00000 V, 10.0000 V, 100.000 V, 750.00 V. 5. For 750 V range accuracy is specified for input less than 200 Vrms. 6. For 100 mV range, the 10 kHz to 30 kHz accuracy is 1.5 + 0.3. 7. Additional error 0.003% of full scale per kHz to be added when signal input changes less than 10% of range. 8. For 100 mV range, the 30 kHz to 100 kHz accuracy is 5 + 0.3. 9. For 750 V range accuracy is specified for input less than 300 Vrms. 10. Available ranges: 10.0000 mA, 100.000 mA, 1.00000 A, 3.0000 A. 11. For 1 A and 3 A ranges, the accuracy is specified for frequencies less than 5 kHz. 5 Digital multimeter specifications (continued) Frequency specifications Frequency range: minimum input frequency = 1 Hz Accuracy (% of reading + % of range) 1 year (23 C 5 C) Temperature coefficient 0 C to 18 C 28 C to 55 C Function Input range Frequency Voltage: 100 mV to 750 V1 < 2 Hz < 20 Hz 20 Hz to 100 kHz 100 kHz to 300 kHz 0.18 + 0.003 0.04 + 0.003 0.02 + 0.003 0.02 + 0.003 0.005 0.005 0.005 0.005 Current: 10 mA to 3 A < 2 Hz < 20 Hz 20 Hz to 10 kHz 0.18 + 0.003 0.04 + 0.003 0.02 + 0.003 0.005 0.005 0.005 Input range Minimum frequency sensitivity for voltage measurement (RMS sine wave) 20 Hz to 100 kHz 100 kHz to 300 kHz 300 kHz to 1 MHz 100 mV 1.0 V 10 V 100 V 750 V 50 mV 100 mV 1V 10 V 100 V Input range Minimum frequency sensitivity for current measurement (RMS sine wave) 20 Hz to 100 kHz 10 mA 100 mA 1A 3A 1 mA 10 mA 100 mA 300 mA 2 50 mV 120 mV 1.2 V 12 V - 0.5 V 0.5 V - - - 1. For 100 mV and 1 V ranges, the measurable frequency is up to 1 MHz at 0.5 V input signal. 2. Maximum input for the specified accuracy = 10 x range or 1000 Vdc. All frequency counters are susceptible to errors when measuring low-voltage, low-frequency signals. Shielding inputs from external noise pickup is critical for minimizing measurement errors. Duty cycle and pulse width specifications Function Range Resolution Accuracy at full scale Duty cycle 100.000%1 0.001% 0.3% + 0.2% per kHz 0.001 ms 0.01 ms Duty cycle/frequency Duty cycle/frequency Pulse width 2 199.999 ms 1999.99 ms2 1. The range is from {10 s x frequency x 100%} ~ {[1 - (10 s x frequency)] x 100%}. For example, a 1 kHz signal can be measured from 1% ~ 99%. 2. The positive or negative pulse width must be greater than 10 s. The range of the pulse width is determined by the frequency of the signal. 6 Digital multimeter specifications (continued) Measurement speeds (typical) Reading speed over GPIB3 (readings/second) Function Rate Reading speed (readings/second) Reading speed over USB2 (readings/second) DC voltage (10 V) Slow (51/2 digits) Fast (41/2 digits) 15 70 17 31 17 32 DC current (1 A) Slow (51/2 digits) Fast (41/2 digits) 15 70 17 37 17 36 AC voltage (10 V at 1 kHz) Slow (51/2 digits) Fast (41/2 digits) 15 70 17 31 17 32 AC current (1 A at 1 kHz) Slow (51/2 digits) Fast (41/2 digits) 15 70 16 37 17 37 Resistance (100 k) Slow (51/2 digits) Fast (41/2 digits) 15 70 17 27 17 31 Capacitance (10 F) Slow/Fast (31/2 digits) 5 4.4 4.6 Frequency (voltage path at 10 V, 1 kHz) Slow (51/2 digits) Fast (41/2 digits) 9 9 2.7 2.7 2.7 2.7 Frequency (current path at 10 V, 1 kHz) Slow (51/2 digit) Fast (41/2 digit) 9 9 2.7 2.7 2.7 2.7 1 1. Reading rate of the A/D converter. 2. Number of measurements per second that can be read through USB using SCPI "READ?" command. 3. Number of measurements per second that can be read through GPIB using SCPI "READ?" command. Supplementary specifications DC voltage Measurement method Sigma Delta A-to-D converter Maximum input voltage 1000 Vdc on all ranges Input impedance 10 M 2% range (typical) in parallel with capacitance < 120 pF Input protection 1000 Vrms on all ranges, < 0.3 A short circuit Response time Approximately 0.15 s when the displayed reading reaches 99.9% DC value of the tested input signal at the same range DC current Measurement method Sigma Delta A-to-D converter Maximum input current 10 mA to 3.0 A DC Burden voltage and shunt resistance * * * * Input protection Protected with 3.15 A/500 V, FF fuse Response time Approximately 0.15 s when the displayed reading reaches 99.9% DC value of the tested input signal at the same range < 0.2 V, 10 for 10 mA range < 0.2 V, 1 for 100 mA range < 0.3 V, 0.1 for 1 A range < 0.7 V, 0.01 for 3 A range 7 Digital multimeter specifications (continued) Supplementary specifications AC voltage Measurement method AC coupled true rms Maximum input voltage 750 Vrms/1200 Vpeak/3 x 107 V-Hz of product Input impedance 1 M 2% range (typical) in parallel with capacitance < 120 pF Input protection 750 Vrms on all ranges Crest factor For < 5:1 errors included. Limited by the peak input and 100 kHz bandwidth. Maximum 3.0 at full scale. Peak input 300% of range. Limited by maximum input. Response time Approximately 2.5 s when the displayed reading reaches 99.9% AC rms value of the tested input signal at the same range. Overload ranging Will select higher range if peak input overload is detected during auto range. Overload is reported in manual ranging. AC current Measurement method AC coupled true rms Maximum input current 10 mA to 3.0 A DC or AC rms Burden voltage and shunt resistance * * * * Input protection Protected with 3.15 A/500 V, FF fuse Crest factor For < 5:1 errors included. Limited by the peak input and 100 kHz bandwidth. Maximum 3.0 at full scale. Peak input 300% of range. Limited by maximum input. Response time Approximately 2.5 s when the displayed reading reaches 99.9% AC rms value of the tested input signal at the same range. < 0.2 V, 10 for 10 mA range < 0.2 V, 1 for 100 mA range < 0.3 V, 0.1 for 1 A range < 0.7 V, 0.01 for 3 A range Resistance Measurement method Two-wire, open-circuit voltage limited to < 5 V Open circuit voltage < +5.0 Vdc Input protection 1000 Vrms on all ranges, < 0.3 A short circuit Response time Approximately 0.15 seconds for 1 M and ranges below 1 M Low-resistance Measurement method Four-wire; test current is sent from the FORCE terminals and resistance measured at the SENSE terminals. Input protection * FORCE terminals: 3.15 A/250 V * SENSE terminals: 1000 Vrms on all ranges, < 0.3 A short circuit Continuity Measurement method 0.83 mA 0.2% constant current source Open circuit voltage < +5.0 Vdc Audible tone Continuous beeping when reading is less than the threshold resistance of 10 at 1.0 k range Input protection 1000 Vrms on all ranges, < 0.3 A short circuit 8 Digital multimeter specifications (continued) Supplementary specifications Diode Measurement method 0.83 mA 0.2% constant current source Open circuit voltage < +5.0 Vdc Audible tone * Continuous beeping when level is below +50 mV DC * Single tone for normal forward-biased diode or semiconductor junction where 0.3 V reading 0.8 V Input protection 1000 Vrms on all ranges, < 0.3 A short circuit Capacitance Measurement method Computed from constant current source charge time, typical 0.2 V to 1.4 Vac signal level Maximum voltage at full scale * For 1 nF to 10 F range: < 1.5 V * For 100 F to 10000 F: 0.33 V Input protection 1000 Vrms on all ranges, < 0.3 A short circuit Response time Approximately 1 s for 100 F and ranges below 100 F Charge and discharge voltage 5 Vpp (approximately from +3 V to -2 V) Frequency Measurement method Reciprocal counting technique Signal level 0.2 V to 1.4 V Input protection * Voltage path: 1000 Vrms on all ranges, < 0.3 A short circuit * Current path: Protected with 3.15 A/500 V, FF fuse Maximum display counts (excluding frequency) 51/2 digits 120,000 41/2 digits 12,000 Noise rejection Common mode rejection ratio (CMRR) for 1 k unbalanced in LO lead * DC: 140 dB * AC: 70 dB Normal mode rejection ratio (NMRR) * 60 Hz 0.1%: 51/2 digits: 65 dB, 41/2 digits: 0 dB * 50 Hz 0.1%: 51/2 digits: 55 dB, 41/2 digits: 0 dB 9 Power supply specifications Specification assumptions: * * * * * Specifications stated are after 60-minutes of warm-up with no load Operating temperature: 18 C to 28 C (64.4 F to 82.4 F) Accuracy is expressed as (% of output + offset) at 23 C 5 C Temperature coefficient: Add [0.1 x (the specified accuracy) / C] for 0 C to 18 C and 28 C to 55 C Relative humidity (RH) up to 80% at 30 C, proportional to 50% for 30 C to 55 C DC power supply specifications Output ratings * Range S1: 0 V to 30 V, 0 A to 1 A * Range S2: 0 V to 8 V, 0 A to 3 A Programming accuracy 1 year (@ 23 C 5 C), (% of output + offset) * 0.05% + 5 mV * 0.15% + 3 mA Readback accuracy 1 year over GPIB and USB or front panel with respect to actual output (@ 23 C 5 C), (% of output + offset) * 0.05% + 5 mV * 0.15% + 3 mA Ripple and noise With outputs ungrounded, or with either output terminal grounded, 20 Hz to 1 MHz3 * < 2 mVrms; < 30 mVpp * < 1 mArms Front terminal load regulation (% of output + offset) * < 3 mV1 * < 0.03% + 0.3 mA Rear terminal load regulation (% of output + offset) * < 0.01% + 3 mV * < 0.03% + 0.3 mA Line regulation Voltage: 3 mV Current: 1.5 mA Programming resolution 1 mV, 0.1 mA Readback resolution 1 mV, 0.1 mA Front panel resolution 1 mV, 0.1 mA Transient response time Less than 450 ms for output to recover to within 50 mV following a change in output current from full load to half load or vice versa Command processing time Average time for output voltage to begin to change after receipt of digital data when instrument is connected directly to the USB or GPIB is less than 100 ms Overvoltage protection (for CC mode) * Accuracy: 0.5% + 0.5 V * Activation time2: < 2ms Overcurrent protection (for CV mode) * Accuracy: 0.5% + 0.05 A * Activation time2: < 2 ms 1. Contacts and leads resistance may contribute an additional error of 6 mV/A (typical). 2. Average time for the detection of OVP or OCP condition. 3. Refer to U3606A User's and Service Guide (U3606-90013); CV noise effect measurement on page 166. 10 Power supply specifications (continued) Sweep specifications Function Scan Range Constant voltage S1 (30 V/1 A) S2 (8 V/3 A) Constant current S1 (30 V/1 A) S2 (8 V/3 A) Ramp Constant voltage S1 (30 V/1 A) S2 (8 V/3 A) Constant current S1 (30 V/1 A) S2 (8 V/3 A) 1. Amplitude start position is fixed at 0 (V or A) by default. Amplitude1 Step Dwelling time 0 to 31.500 V 0 to 8.4000 V 0 to 1.0500 A 0 to 3.1500 A 0 to 31.500 V 0 to 8.4000 V 0 to 1.0500 A 0 to 3.1500 A 1 step to 100 steps 1 s to 99 s 1 step to 100 steps 1 s to 99 s 1 step to 10,000 steps 300 ms/step (typical) 1 step to 10,000 steps 300 ms/step (typical) Square-wave output specifications Parameter Range Frequency 0.5, 2, 5, 6, 10, 15, 25, 30, 40, 50, 60, 75, 80, 100, 120, 150, 200, 240, 300, 400, 480, 600, 800, 1200, 1600, 2400, 4800 Hz Duty cycle 0.39% to 99.60% Pulse width 1/frequency Amplitude S1 (30 V/1 A) S2 (8 V/3 A) 1. Specification applies when the positive or negative pulse width is greater than 50 s. Resolution Accuracy 0.01 Hz 0.005% + 1 count 0.39%1 Range/256 1 mV 1 mV 0.4%1, 2 Duty cycle/frequency1, 3 0.2 V 0.2 V 2. For frequency signals greater than 100 Hz, an additional 0.1% per 100 Hz is to be added. Accuracy = 0.4% + - 1 ) x 0.1% ( frequency 100 3. The accuracy of pulse width could also be calculated as [0.4% + (frequency/100 - 1) x 0.1%]/frequency. Rise/fall time is less than 25 s. Specifications are based on a resistive load. Supplementary specifications Output programming range Range Output programming OV/OC OVP/OCP S1 S2 31.500 V/1.05 A 8.4 V/3.15 A 33.000 V/1.1 A 8.8 V/3.3 A 0 V to 31.500 V/0 A to 1.05 A 0 V to 8.4 V/0 A to 3.15 A Remote sensing capability Range S1 (30 V/1 A) Range S2 (8 V/3 A) Up to a 0.75-volt drop per load lead Up to a 0.5-volt drop per load lead Temperature coefficient: (% of output + offset)/ C for 0 C to 18 C and 28 C to 55 C Voltage Current 0.005% + 0.5 mV/C 0.02% + 1 mA/C Voltage programming speed (excludes command processing time) Up Down 300 ms (full load and no load) 400 ms (full load and no load) 11 General characteristics Power supply * Universal 100 Vac to 240 Vac 10% * AC line frequency 45 Hz to 66 Hz (360 Hz to 440 Hz for 100/120 V operation) Power consumption 150 VA maximum Current Input Fuse 3.15 A, 500 V FF fuse (on front panel) Display Highly visible vacuum-fluorescent display (VFD) Operating environment * * * * * Operating temperature from 0 C to +55 C Relative humidity up to 80% at 30 C RH (non-condensing) Altitude up to 2000 meters Pollution degree 2 For indoor use only Storage compliance -40 C to 70 C Safety compliance Certified with: * IEC 61010-1:2001/EN61010-1:2001 (2nd Edition) * Canada: CAN/CSA-C22.2 No. 61010-1-04 * USA: ANSI/UL 61010-1:2004 EMC compliance Certified with: * IEC61326-1:2005 / EN61326-1:2006 * CISPR 11:2003 / EN55011:2007 (Group 1 Class A) * Canada: ICES/NMB-001:2004 * Australia/New Zealand: AS/NZS CISPR11:2004 Shock and vibration Tested to IEC/EN 60068-2 Remote interface * * * * Programming language Standard Commands for Programmable Instruments (SCPI) Measurement category * CAT II, 300 V * CAT I, 1000 Vdc, 750 Vac rms * 2500 Vpk transient over voltages Dimensions (H x W x D) 105 x 255 x 329 mm (with rubber bumpers) 87 x 215 x 312 mm (without rubber bumpers) Weight 3.775 kg (with rubber bumpers) 3.535 kg (without rubber bumpers) Warranty Three years Calibration cycle One year Warm-up time 60 minutes GPIB IEEE-488 Full Speed USB 2.0 (Type B) USBTMC 488.2 Class device USB-CDC 12 Ordering information Standard shipped items Warranty options I/O connectivity options * * * * * * * R-51B-001-3C Extended warranty from one year to three years * R-51B-001-5C Extended warranty from one year to five years For control via GPIB interface * 82350B/82351A PCI/PCIe high-performance GPIB interface card * 82357B USB/GPIB converter * E5810A LAN/GPIB gateway * 10833D/A/B/C/F/G GPIB cables * 10834A GPIB-to-GPIB adapter Quick Start Guide Product Reference CD Agilent IO Library Suite Certificate of Calibration U8201A Combo Test Lead Kit USB 2.0 High-Speed Type-A to Type-B cable * AC power cord For control via USB interface * E5813A networked 5-port USB hub Optional accessories U8201A Combo Test Lead Kit U3606A-1CM Rack Mount Kit U8202A Electronic Test Lead Kit (for DMM function) 34133A Precision Electronic Test Leads (for DMM function) 34330A Current Shunt (30 A) (for DMM function) 34136A 40 kV high-voltage probe (for DMM function) 11059A Kelvin Probe Set and 11062A Kelvin Clip Set (for DMM function) E3600A-100 Test Lead Kit (for DC power supply function) 13 www.agilent.com www.agilent.com/ind/U3606A myAgilent myAgilent www.agilent.com/find/myagilent A personalized view into the information most relevant to you. For more information on Agilent Technologies' products, applications or services, please contact your local Agilent office. The complete list is available at: www.agilent.com/find/contactus www.axiestandard.org AdvancedTCA(R) Extensions for Instrumentation and Test (AXIe) is an open standard that extends the AdvancedTCA for general purpose and semiconductor test. Agilent is a founding member of the AXIe consortium. www.lxistandard.org LAN eXtensions for Instruments puts the power of Ethernet and the Web inside your test systems. Agilent is a founding member of the LXI consortium. www.pxisa.org PCI eXtensions for Instrumentation (PXI) modular instrumentation delivers a rugged, PC-based high-performance measurement and automation system. Three-Year Warranty www.agilent.com/find/ThreeYearWarranty Beyond product specification, changing the ownership experience. Agilent is the only test and measurement company that offers three-year warranty on all instruments, worldwide. Agilent Assurance Plans www.agilent.com/find/AssurancePlans Five years of protection and no budgetary surprises to ensure your instruments are operating to specifications and you can continually rely on accurate measurements. www.agilent.com/quality Agilent Electronic Measurement Group DEKRA Certified ISO 9001:2008 Quality Management System Americas Canada Brazil Mexico United States (877) 894 4414 (11) 4197 3600 01800 5064 800 (800) 829 4444 Asia Pacific Australia China Hong Kong India Japan Korea Malaysia Singapore Taiwan Other AP Countries 1 800 629 485 800 810 0189 800 938 693 1 800 112 929 0120 (421) 345 080 769 0800 1 800 888 848 1 800 375 8100 0800 047 866 (65) 375 8100 Europe & Middle East Belgium Denmark Finland France Germany Ireland Israel Italy Netherlands Spain Sweden United Kingdom 32 (0) 2 404 93 40 45 45 80 12 15 358 (0) 10 855 2100 0825 010 700* *0.125 /minute 49 (0) 7031 464 6333 1890 924 204 972-3-9288-504/544 39 02 92 60 8484 31 (0) 20 547 2111 34 (91) 631 3300 0200-88 22 55 44 (0) 118 927 6201 Agilent Channel Partners For other unlisted countries: www.agilent.com/find/channelpartners www.agilent.com/find/contactus Get the best of both worlds: Agilent's measurement expertise and product breadth, combined with channel partner convenience. (BP-09-27-13) Product specifications and descriptions in this document subject to change without notice. (c) Agilent Technologies, Inc. 2011, 2014 Published in USA, January 10, 2014 5990-3971EN