63W Medical Grade Desktop Switching Power Adapters AMP6301 series
WWW.ASTRODYNE.COM ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
Universal 90-264VAC Input
UL/cUL60601-1, EN60601-1
Regulated Single Outputs
4000V Isolation
IEC320-C14 Input
LPS Efciency Level V (9-36 Vout)
Model Number Output Voltage Output Amps Regulation Max. Output Watts Output Connector
SINGLE OUTPUT
AMP6301-02 5 9 7% 45W DIN
AMP6301-03 7 7.85 7% 55W DIN
AMP6301-04 9 6.44 5% 58W DIN
AMP6301-05 12 5.25 5% 63W DIN
AMP6301-06 15 4.2 5% 63W DIN
AMP6301-07 18 3.5 5% 63W DIN
AMP6301-08 24 2.62 3% 63W DIN
AMP6301-09 30 2.1 3% 63W DIN
AMP6301-10 36 1.75 3% 63W DIN
2.99”W x 5.75”L x1.69”H
63W Medical Grade Desktop Switching Power Adapters AMP6301 series
WWW.ASTRODYNE.COM ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
INPUT SPECIFICATIONS
Input Voltage Range 90-264 VAC
Frequency Range 47-63 Hz
Leakage Current 0.1mA typ. @ 240VAC
Input Current Lowline 1.62A
Input Current Highline 0.72A
Inrush Current Lowline 15A max.
Inrush Current Highline 30A max.
Power Consumption @ No Load (Approved to CEC level IV)
12-36 Vout: 0.5W, max.
OUTPUT SPECIFICATIONS
Voltage and Current See Selection Charts
Load Regulation See Selection Charts
Line Regulation ± 0.5%, typ (1% Max.)
Ripple/Noise 0.5% Pk-Pk, typ (1% Max.)
Hold Up Time 16mS min. at 110VAC Input
Start Up Time 1S typ. (2S max.)
Over Voltage Protection 112-132%
Over Current Protection 110-150%
Transient Response (FL-1/2L) 4mS
Temperature Coefcient ±0.04%/°C, typ.
GENERAL SPECIFICATIONS
Isolation I/P-O/P: 5600VDC
I/P-GND: 2800VDC
Isolation Resistance 50MΩ
Efciency 85% typ.
Safety UL60601-1, EN60601-1
PHYSICAL SPECIFICATIONS
Size 5.75”L x 2.99”W x 1.69”H
Construction Desktop Style
Weight 19.75oz (560g)
ENVIRONMENTAL SPECIFICATIONS
Oper. Temperature 0 to +70°C
See Derate Curve
Storage Temperature -40 to +85°C *
Relative Humidity 5 to + 95%, non-cond *
MTBF: MIL-HDBK-217F 100,000 Hrs.
EMI 220VAC in: CISPR-11B
110VAC in: FCC-18B
Efciency Level V (9-36 Vout only)
OUTPUT DERATING CURVE
* These are stress ratings. Exposure of the devices to any of these conditions
may adversely affect long term reliability. Proper operation under conditions
other than the standard operating conditions is neither warranteed nor implied.
All specications are typical at nominal input, full load, and 25°C unless other-
wise noted Astrodyne products are not authorized or warranteed for use
as critical components in life support systems, equipment used
in hazardous environments, nuclear controls systems, or other
mission-critical applications.
63W Medical Grade Desktop Switching Power Adapters AMP6301 series
WWW.ASTRODYNE.COM ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
MECHANICAL DIMENSIONS
DC Output Connections
DIN (Multi-Output Models)
1
2
3
45
5 Pin DIN
1. COM
2. COM
3. Output
4. COM
5. Output
SHIELD = GND
AMP6301-02~07 Output Cord Length 48”, 16AWG
AMP6301-08~10 Output Cord Length 72”, 18AWG