Amprobe® Test Tools www.Amprobe.com
Data Sheet
Electrical Specifications
Temperature Scale Celsius or Fahrenheit user-selectable
Measurement Range K-Type -50°C to 1000°C, (-50°F to 1832°F)
Resolution 0.1°C or 0.2°F
Accuracy Accuracy is specified for operating temperatures over the range of 18°C to 28°C (64°F to 82°F), for 1
year, not including thermocouple error. ±(0.1% rdg + 1°C) on -50°C to 1000°C ±(0.1% rdg + 2°F) on
-50°F to 1832°F ±(0.05% rdg + 1.4°F) -58°F to -328°F
Sensor Thermistor temperature sensor
Range 0°C to 60°C, (32°F to 140°F)
Resolution 0.1°C/°F
Accuracy ±2°C on 0°C to 10°C ±0.5°C on 10°C to 45°C ±2°C on 45°C to 60°C ±4°F on 32°F to 50°F ±1°F on 50°F to
113°F ±4°F on 113°F to 140°F
RELATIVE HUMIDITY
Sensor Capacitive Humidity Sensor
Range 0% to 100% RH
Accuracy ±2.5% at 25°C (77°F), 10% to 90% RH,
±5% at 25°C (77°F), 0% to 10% RH,
90% to 100% RH
Sensor Response Time 60 sec typical
for 90% of Total Range
Sensor Hysteresis ±1%RH typical
(excursion of 10% to 90% to 10% RH)
Temperature 0.1 times the applicable accuracy
Coefficient Specification per °C from 0°C to 18°C and 28°C to 50°C (32°F to 64°F and 82°F to 122°F).
Input Protection 24V dc or 24V ac rms maximum input Voltage on any combination of input pins.
Input Connector Accepts standard miniature thermocouple connectors (flat blades spaced 7.9mm, center to center).
THWD-10W RH Meter with Wireless