2700 SERIES
Table 1: Absolute Maximum Ratings
Parameter Symbol Test Conditions Minimum Maximum
Operating case temperature Top - -5 °C 75 °C
Storage temperature Tstg 2000 hours -40 °C 85 °C
LD submount temperature TLD - 0 °C 50 °C
LD reverse voltage Vr- - 2 V
LD forward current - - 650 mA
LD current transient 1 µs maximum - 1000 mA
LD reverse current - - 10 µA
PD reverse voltage VPD - - 20 V
PD forward current IPD - - 10 mA
Electrostatic discharge (ESD) VESD C = 100 pF, R = 1.5 Ω, human body model - 1000 V
TEC current ITEC - - 2.5 A
TEC voltage VTEC - - 4.0 V
Axial pull force 3 x 10 seconds - 5 N
Side pull force 3 x 10 seconds - 2.5 N
Fiber bend radius - 16 mm -
Relative humidity RH40 °C 5% 95%
Lead soldering time 260 °C - 10 seconds
Note: Absolute maximum ratings are the maximum stresses that may be applied to the pump module for short periods of time without causing damage. Stresses in excess of the absolute maximum ratings can
permanently damage the device. Exposure to absolute maximum ratings for extended periods, or exposure to more than one absolute maximum rating simultaneously may adversely affect device reliability.
Table 2: Operating Parameters
Product Maximum Maximum Minimum Kink-Free
Code1Operating Operating Kink-Free Current
Power Current Power Imax (mA)3
Pop (mW)2,3 Iop (mA) Pmax (mW)4Maximum4
Maximum2
27-xxxx-190 170 360 190 400
27-xxxx-200 180 380 200 420
27-xxxx-210 190 400 210 440
27-xxxx-220 200 420 220 460
27-xxxx-230 210 440 230 480
27-xxxx-240 220 460 240 500
27-xxxx-250 230 480 250 520
27-xxxx-260 240 500 260 540
27-xxxx-270 250 520 270 560
27-xxxx-280 260 540 280 580
27-xxxx-290 270 560 290 600
27-xxxx-300 280 560 300 600
1. The 29-xxxx-xxx-FL may be substituted for this part series. All 27-xxxx-xxx end product specifications will remain as published; there is no change to part numbers, product testing, quality or reliability.
Traceability is accomplished through product serial number.
2. The maximum operating power Pop will be achieved at a device-specific current, the maximum operating current Iop. The individual value of Iop is noted on the hardcopy data sheet shipped with the device.
All values of Iop are limited by the maximum value listed in Table 2.
3. The pump laser shall never be operated at a power higher than the maximum operating power Pop throughout its lifetime. At Begin of Life (BOL), the operating current shall never be higher than the
device-specific maximum operating current Iop that is noted on the hardcopy data sheet shipped with the device. At End of Life (EOL), the operating current shall never be higher than the device-specific
kink-free current Imax that is noted on the hardcopy data sheet shipped with the device.
4. The module is kink-free (at least) up to a minimum kink-free power Pmax that the module will achieve at a device-specific current, the kink-free current Imax. The individual value of Imax is noted on the
hardcopy data sheet shipped with the device. All values of Imax are limited by the maximum value listed in Table 2.
(BOL, Tcase = 0 to 75 °C, TLD = 25 °C, -50 dB reflection, unless noted otherwise)